Physiological Neonatal Resuscitation at Birth

Hutchon DJ R

Published on: 2019-09-10

Abstract

About 10% of term neonates are transferred to the room side resuscitation equipment when positive pressure ventilation [PPV] is thought to be required. About one third of the neonates transferred start breathing sufficiently before PPV can be provided so that no active care is necessary. Transfer to the room side resuscitation equipment can only be carried out after immediate cord clamping. Transition after early cord clamping is not physiological and cord clamping abruptly removes the placental circulation before the pulmonary circulation has been established which disrupts a smooth transition. Those neonates who are truly asphyxiated benefit from the intervention of positive pressure ventilation but all the babies removed from their mother for room side ventilation will have suffered the consequences of early cord clamping. Reversal of the asphyxia will only occur after a good respiratory exchange has taken place within the lungs. In addition to expansion of the lungs and establishment of the functional residual capacity an adequate circulation is essential. However the marked increase in afterload of the heart and the reduction in preload and cardiac output caused by early cord clamping may reduce the pulmonary circulation. Thus the condition of the asphyxiated neonate who is not ventilated until after the cord has been clamped will be further compromised by cord clamping. Cord clamping is an intervention with no proof of benefit in the apnoeic neonate apparently requiring resuscitation. It might be considered a pragmatic solution in the asphyxiated neonate, as there is limited experience of resuscitation at the mother’s side with an intact cord. However, from the limited experience that there is, we do know that resuscitation can be carried out effectively with an intact placental circulation by the side of the mother. The procedures need to be clearly understood, practiced and co-ordinated by the whole team of obstetricians, neonatologists and midwives before the clinical benefits for the neonate can be realised. This paper will summarise the physiological arguments for resuscitation with an intact cord, and provide a range of novel procedures and equipment which can avoid the further intervention of cord clamping and mother side neonatal resuscitation with an intact cord. Thus the benefits of delayed or physiological cord clamping will be available to all neonates at all modes of birth.

Keywords

Physiological Transition; Neonatal Resuscitation; Asphyxia; Hypovolaemia; Hypoxia

Introduction

About 10% of term neonates are transferred to the room side resuscitation equipment when positive pressure ventilation [PPV] is thought to be required [1]. About one third of these neonates start breathing sufficiently before PPV can be provided so that no active care is necessary. Transfer of the apnoeic neonate to the room side resuscitation equipment can only be carried out after immediate cord clamping. Cord clamping before inflation of the lungs is established is not physiological and disrupts the transition from placental to pulmonary respiration [2]. In the asphyxiated neonate this disruption can potentially be avoided by initiating PPV before the cord is clamped [3]. This requires resuscitation with an intact cord at the side of the mother.

Rationale and evidence base for mother side resuscitation with an intact cord

Early cord clamping before the onset of breathing is not physiological and the intervention adversely affects transition from placental to pulmonary respiration [4]. A delay of at least one minute after birth is recommended in all major guidelines [1,5]. The most widely confirmed effect is the loss of the placental transfusion and loss of this blood volume results in a variable degree of hypovolaemia. The result of this blood loss is anaemia and iron deficiency in the following months, and the iron deficiency has been linked to impaired neurodevelopment [7-11]. For many years bradycardia has been known to follow early cord clamping [12] and more recently bradycardia and a marked fall in cardiac output in animal studies has been demonstrated [2]. In preterm neonates mortality and the use of blood transfusion are reduced if early cord clamping is avoided [13]. Other studies have shown a reduction in intra ventricular haemorrhage, necrotising enter colitis and late onset sepsis [14]. While there is plenty of evidence to show the harm of early cord clamping in healthy neonates there is also sufficient evidence to support a change in practice for asphyxiated neonates [15,16]. The loss of the placental transfusion can be over 200mls [6]. Dunn and others warned of the risk of severe hypovolaemia with early cord clamping and intra partum cord clamping [17]. Neonates with severe hypovolaemia may not respond to ventilation until the hypovolaemia is corrected with an emergency transfusion of O-nag blood [18]. Outcomes are improved in neonates with congenital heart abnormality [19] and also in neonates with congenital diaphragmatic hernia [20] if early cord clamping is avoided and a physiological transition permitted.
Extrapolating this evidence suggest that providing mother side resuscitation with an intact cord in asphyxiated neonates should result in less morbidity and mortality. Asphyxia is not a single entity but a spectrum from mild to severe and may not be accurately predicted or assessed during the first seconds after birth [21]. If mother side resuscitation is available, neonates who are falsely identified as asphyxiated and in whom this cord clamped immediately to move them to a room side resuscitation trolley, will be spared the harm of early clamping. Feasibility studies have shown that the approach of mother side resuscitation is practical and achievable [22-27].

Practical provision of mother side resuscitation after normal birth

Normal birth presents the least difficulty in providing mother side neonatal resuscitation with an intact cord. The neonate can remain close to the perineum under one of the mother’s leg [7]. All standards of care a such as open airway and thermoregulation of the neonate during the recommended delay of at least one minute can be provided in the normal routine manner and PPV can be provided by a self-inflating bag and mask when necessary. This level of care is provided by some midwives [26]. These are low risk neonates but there remains a small risk of apnoea at birth and preparation is required in anticipation of the small risk of apnoea.

Practical provision of mother side resuscitation at caesarean section

The temperature of the delivery room is very important to reduce the risk of neonatal hypothermia. It must be well above the normal operating theatre temperature and more aligned with the ambient temperature of the neonatal nursery. The need to ensure that the sterile field is not compromised is a major challenge in providing mother side resuscitation at caesarean section. In addition there is no obvious surface on which to place the neonate and still meet the standard of care for an open airway. A customised solution is required. A modified standard resuscitation trolley can be brought right up to the side operating table but this is cumbersome and may interfere with the position of the surgical team [28]. Since the height of the trolley is fixed and the height of the operating table needs to be adjusted to match the resuscitation trolley. The Life Start is a much smaller customised trolley with a variable height and includes all the normal ventilation equipment [usually PEEP with warmed humidified air] and thermal equipment. The resuscitation platform is covered with a sterile towel to maintain sterility [29]. The Concord is another customised design which incorporates a long extendable platform for the neonate, allowing it to remain close enough to its mother to provide resuscitation with an intact cord [30]. In the Vent first trial, a large multicentre trial of mother side resuscitation in preterm neonates, all the resuscitation equipment is mounted on a mobile pole. A platform for the neonate and warming arrangements are customised locally [31].

Mother side resuscitation at assisted vaginal birth

The same customised trolley for caesarean birth can be used at an assisted vaginal birth if the platform for the neonate can be low enough to allow the resuscitation team to reach the neonate below the mother’s legs in lithotomy poles. The height adjustable platform is a design feature of the Life Start and Concord trollies.

Training and simulation practice

Such a marked change in practice cannot be undertaken incrementally as has been possible with some other changes in neonatal and obstetric practice. It requires close team work between the two separate disciplines of obstetrics and neonatology. The co-operation needs to be provided during the critical few minutes after birth, often with little notice. The high standard of care which both team members expect to be able to provide cannot be met at mother side neonatal resuscitation unless the procedures and responsibility have carefully prepared and practiced. This is probably just as important as the design of the customised equipment necessary to provide a level of neonatal care and resuscitation normally available at room side resuscitation. When mother side resuscitation is first introduced to a unit it is essential that there is initial training, simulation and assessment [32]. A locally prepared protocol must be available. After each case the equipment and procedures should be reviewed so that any difficulties can be resolved. It has been shown that performance in stressful resuscitation scenarios Human Factors and Ergonomics [HFE] principles are important. They have been successfully applied to the organization of resuscitation equipment for neonates [33].

Discussion

Active management of the third stage of labour to reduce the risk of postpartum haemorrhage [PPH] originally included the practice of early cord clamping [34]. Early cord clamping was only recently proven to be unnecessary to reduce PPH and as a result early cord clamping has been removed from the recommendations. A delay of at least one minute is recommended in healthy neonates and ventilation is considered the priority in neonates requiring resuscitation. [5]. Ventilation can be provided with an intact cord if the skill and experience is available to provide effective ventilation with an intact cord [35]. To make neonatal resuscitation more in line with physiology, early cord clamping must be avoided [36]. There are many reasons however why mother side resuscitation with an intact cord is likely to be much more effective and safer. It is very much less likely to be associated with neonatal hypovolaemia which occurs with early cord clamping especially severe after intra partum cord compression. In spite of apparently responding well to resuscitation, infants in poor condition at birth have an increased risk of poor functioning in cognitive tests in later life and this could partly be a result of early cord clamping [37]. With current practice a small number of neonates requiring resuscitation fail to respond to ventilation. In some of these hypovolaemia is the underlying problem [14]. With the redistribution of blood from the placenta to the neonate which occurs within the few minutes after birth and before cord clamping, significant hypovolaemia is much less likely. If prolonged ventilation is still required after mother side resuscitation with an intact cord, the cord can be clamped and cut when the placental circulation has ceased and the neonate removed to room side resuscitation equipment. The transition of the circulation will have been close to a physiological transition [3] Parents have accepted removal of their new born baby immediately after birth in the belief that this benefited their baby. Resuscitation with relatives present in other situations is now common. Having the parents so close and seeing every intervention on their new born may put pressure on the carers, or frighten the parents. However the evidence shows that the carers appreciate the close presence of the parents as much as the parents appreciate remaining close to their new born [38]. Neonates may also respond better to resuscitation when they can hear the parents’ voices or feel their touch [39]. Clearly the placental circulation must be intact for the procedure to be beneficial and there are rare contraindications to the approach. Significant bleeding from a vasa Previa prevents an adequate placental circulation. A short cord may prevent the neonate from reaching the resuscitation platform, however with experience it may be possible to get the neonatal platform sufficiently close to the mother even with a short cord. Cord milking may avoid the loss of the placental transfusion but still results in cord clamping before expansion of the lungs and a physiological transition, and clinical evidence is lacking.

References

  1. Wyllie J, Perlman J M, Kattwinkel J, Wyckoff MH, Aziz K, Guinsburg R, et al. neonatal resuscitation, international consensus on cardiopulmonary resuscitation and emergency, cardiovascular care science with treatment recommendations, Resuscitation.2015; 95:169–201.
  2. Bhatt S, Alison BJ, Wallace EM, Crossley KJ, Gill AW, Kluckow M, et al. Delaying cord clamping until ventilation onset improves cardiovascular function at birth in preterm lambs, J Physiol . 2013; 591:2113–2126.
  3. Martinello K, Hart AR, Yap S, Mitra S, Robertson N. Management and investigation of neonatal encephalopathy, update Arch Dis Child Fetal Neonatal Ed. 2017; 102:346–358.
  4. Polglase, G R, Dawson, J A , Kluckow, M. et al, Ventilation onset prior to umbilical cord clamping (physiological-based cord clamping) improves systemic and cerebral oxygenation in preterm lambs. 2015; 10:117-504.
  5. National Institute for Health and Care Excellence, Intrapartum care,  NICE guideline CG190. 2014;
  6. Farrar D, Airey R, Law G, Tuffnell D, Cattle B, Duley L, et al. Measuring placental transfusion for term Births, weighing babies with cord intact BJOG. 2011; 118:70–75.
  7. van Rheenen P. Editorial Delayed cord clamping and improved infant outcomes, BMJ 2011; 343:7127.
  8. Andersson O, Hellström-Westas L, Andersson D, Domellöf M. Effect of delayed versus early umbilical cord clamping on neonatal outcomes and iron status at 4 months, a randomised controlled trial BMJ. 2011; 343:7157.   
  9. Andersson O, Domellöf M, Andersson D, Hellström-Westas L. Effect of delayed vs early umbilical cord clamping on iron status and neurodevelopment at age 12 months a randomized clinical trial, JAMA Pediatr. 2014; 168:547-554.
  10. Ashish, K C, Malqvist, M , Rana, N , Ranneberg, L J , Andersson, O, et al. Effect of timing of umbilical cord clamping on anemia at 8 and 12 months and later neurodevelopment in late pre-term and termn infants, a facility-based, randomized-controlled trial in Nepal, BMC Pediatr. 2016; 16:35
  11. Baenziger O, Stolkin F, Keel M, von Siebenthal K, Fauchere J C, Das Kundu S, et al. The influence of the timing of cord clamping on postnatal cerebral oxygenation in preterm neonates a randomized, controlled trial, Pediatrics. 2007; 9:119-455.
  12. Brady JP, James LS. Heart rate changes in the fetus and newborn infant during labor, delivery and the immediate neonatal period, Am J Obstet Gynecol. 1962; 84:1-12.
  13. Fogarty M, Osborn D A, Askiea L, Seidlera A L, Hunter K, et al Delayed vs early umbilical cord clamping for preterm infants a systematic review and meta-analysis, American Journal of Obsetrics and Gynecol.2017; 10-231.
  14. Mercer, J S , Vohr, B R , McGrath, M M , Padbury, J F , Wallach, M , Oh, W, et al.Delayed cord clamping in very preterm infants reduces the incidence of intraventricular hemorrhage and lateonset sepsis a randomized, controlled trial, Pediatrics. 2006; 117:1235–1242.
  15. Manley, B J , Owen, L S , Hooper, S B , et al. Towards evidence based resuscitation of the newborn infant, Lancet. 2017; 389:1639–1648.
  16. Kluckow M, Hooper S B. Using physiology to guide time to cord clamping, Sem Fetal Neonatal Med. 2105; 20:223 -31.
  17. Dunn P M. Tight nuchal cord and neonatal hypovolaemic shock, Archives of Disease in Childhood. 1988; 63:570– 571.
  18. Rennie J M. Neonatology In Bennet PN, Kenny LC eds. Obstetrics by TEN TEACHERS, (19th Edn). Boca Raton, FL CRC Press. 2011; 286.
  19. C H Backes, H Huang, CL Cua. Early versus delayed umbilical cord clamping in infants with congenital heart disease a pilot, randomized, controlled trial, Journal of Perinatol.2015; 35:826–831.
  20. Lefebvre C, Rakza T, Weslinck N, Vaast P, Houfflin debarge V. et al. Feasibility and safety of intact cord resuscitation in newborn infants with congenital diaphragmatic hernia CDH Resuscitation.2017; 08-233. DOI: https://doi.org/10.1016/j.resuscitation
  21. Golubnitschaja O, Yeghiazaryan K, Cebioglu M. Birth asphyxia as the major complication in newborns moving towards improved individual outcomes by prediction, targeted prevention and tailored medical care, EPMA J. 2011; 2:197–210.  
  22. Hutchon DJR. How to resuscitate the neonate with the cord intact at Caesarean Section, British International Congress of Obstetrics and Gynaecology, London. 2007; 02-28.
  23. Weeks AD, Watt RJP, Hutchon DJR. Innovation in immediate neonatal care development of the Bedside Assessment, Stabilisation and Initial Cardiorespiratory support BASICS trolley. 2014.
  24. Katheria A C, Sorkhi S R, Hassen K, Faksh A, Ghorishi Z, Poeltler D, et al. Acceptability of bedside resuscitation with intact umbilical cord to clinicians and patients’ families in the United States, Front Pediatr. 2018; 6-100.
  25. Schoonakker BDJ, Dorling J, Oddie S. Bedside resuscitation of preterm infants with cord intact is achievable using standard resuscitaire, Oporto, Portugal European Society for Pediatric Research.
  26. Fulton C, Stoll K. Thordarson Bedside resuscitation of newborns with an intact umbilical cord Experiences of midwives from British Columbia, Midwifery. 2016; 34:42–46.
  27. Katheria A, Poeltler D, Durham J. Neonatal resuscitation with an intact cord a randomized clinical trial, J Pediatr. 2016; 178:75–80.
  28. Hutchon DJR. How to resuscitate the neonate with the cord intact at Caesarean Section, British International Congress of Obstetrics and Gynaecology. London. 2007; 02-28.
  29. Hutchon D, Bettles N. Mother side care of the term neonate at birth Maternal Health, Neonatol and Perinatol. 2016; 2-5.
  30. Concord Neonatal https://concordneonatal.com/
  31. VentFirst A Multicenter RCT of Assisted Ventilation During Delayed Cord Clamping for Extremely Preterm Infants.
  32. McAdams R M, Backes C H, Hutchon D J R. Steps for implementing delayed cord clamping in a hospital setting, Maternal Health Neonatal Perinatal. 2015; 1-10.
  33. Law B, Cheung P, O’Reilly M, Fray C, Schmölzer G. Reorganizing Neonatal Resuscitation Equipment Improves Performance Speed Under Simulation Conditions, Conference: International Conference on Applied Human Factors and Ergonomics Advances in Intelligent Systems and Computing. 2018; 1-35.  
  34. Prendiville W J, Harding J E, Elbourne D R, Stirrat G M. The Bristol third stage trial active versus physiological management of third stage of labour BMJ. 1988; 297:1295–1300.
  35. WHO, USAID, MCHIP. Delayed Clamping of the Umbilical Cord to Reduce Infant Anaemia. 2013.
  36. McAdams R M, Backes C H, Fathi O, Hutchon D. Revert to the original time to re-establish delayed umbilical cord clamping as the standard approach for preterm neonates Maternal Health, Neonatol and Perinatol.2018; 4-13.
  37. Odd D E , Rasmussen F, Gunnell D, Lewis G, Whitelaw A. A cohort study of low Apgar scores and cognitive outcomes, Arch Dis Child Fetal Neonatal (Ed). 2008; 93:115–120.
  38. Thomas M R, Yoxall C W, Weeks A D. Providing new born resuscitation at the mother’s bedside assessing the safety, usability and acceptability of a mobile trolley. BMC Pediatrics. 2014; 14-135.
  39. Pineda R G, Neil J, Smyser CD. lterations in Brain Structure and Neurodevelopmental Outcome in Preterm Infants Hospitalized in Different Neonatal Intensive Care Unit Environments. 2014; 1: 52-60.